Replacing the air filter:
Replace the air filter every six months (or more often if necessary).
1. Remove the air filter cover from the back of the device.
2. Remove and discard the old air filter.
3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover.
4. Replace the air filter cover.
Notes:
Ensure the air filter and air filter cover are fitted at all times.
Do not wash the air filter. The air filter is not washable or reusable.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your
equipment supplier or ResMed. Do not attempt to open these devices.
Problem/Possible cause
No display
Power is not connected.
The DC plug is partially inserted into the
back of the device or inserted too
slowly.
Insufficient air delivered from the device
Ramp time is in use.
Air filter is dirty.
Air tubing is not connected properly.
Air tubing is blocked, pinched or
punctured.
Mask and headgear are not positioned
correctly.
Incorrect air tubing selected.
4
English
Solution
Ensure the power cable is connected and the power outlet
(if available) is on.
Fully insert the DC plug.
Wait for air pressure to build up or change ramp time.
Replace air filter.
Check air tubing.
Unblock or free the air tubing. Check the air tubing for
punctures.
Adjust position of mask and headgear.
If you are using the SlimLine, Standard or 3m air tubing
ensure that you have the correct air tubing selected via the
menu.