English Indications for use The AutoSet CS-A is indicated to stabilise the ventilation of adult patients exhibiting central sleep apnoea (CSA), mixed sleep apnoea and periodic breathing, with or without obstructive sleep apnoea. It is intended for home and hospital use.
Connect one end of the air tubing firmly onto the air outlet. Connect the assembled mask system to the free end of the air tubing. Note: Always ensure that the AutoSet CS-A is placed in an area where the alarm LED indicators are clearly visible.
Viewing alarms Refer to illustration D. The AutoSet CS-A is fitted with an alarm module that continuously monitors both therapy and device conditions. Alarms are only activated when therapy is running. An alarm condition is indicated by an audible sound, a flashing yellow alarm LED and a message on the screen (refer to D1).
Remove the air filter cover from the back of the device. Remove and discard the old air filter. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover. Replace the air filter cover.
Page 6
Problem/Possible cause Solution The DC plug is partially inserted Fully insert the DC plug. into the back of the device or inserted too slowly. Insufficient air delivered from the device Ramp time is in use. Wait for air pressure to build up or change ramp time. Air filter is dirty.
Page 7
SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system Only use equipment recommended by ResMed. being used. Incorrect mask setting being Check the selected mask type in the Setup menu. Change it if used.
Page 8
Turn the humidity level setting down and empty the water from high, resulting in accumulation the air tubing. of water in the air tubing. Check ResMed 30/90W Power Supply Unit and fully insert the connector Displays message: The DC plug is partially inserted Fully insert the DC plug.
Page 9
Problem/Possible cause Solution The following message is NOT displayed on the LCD after you try to update the settings using Settings updated successfully, press any key the SD card: The settings were not updated. Contact your clinician/service provider immediately. Alarms If the system has not been properly assembled, the device will trigger an alarm.
Page 10
Problem/Possible cause Solution No SpO2 data, check oxi sensor attachment to module/finger Displays message: Oximeter sensor is not attached Ensure that the oximeter sensor is attached properly to the module properly. and the patient's finger. Oximeter sensor might be If the message appears repeatedly but the oximeter is attached faulty.
Page 11
It is recommended that mobile communication devices are kept at least 1 m away from the device. Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on www.resmed.com, on the Products page under Service and Support. Click on the PDF file for your language. IEC 60601-1...
Page 12
AutoSet CS-A technical specifications Mode pressure CPAP mode ranges Set Pressure: 4–20 cm H ASV and ASVAuto mode EPAP: 4–15 cm H O; PS: 0–20 cm H Maximum single Maximum single fault steady state pressure: 30 cm H O—if pressure exceeded fault pressure for >...
Page 13
Air tubing technical specifications Air tubing Material Length Inner diameter ClimateLine heated air tubing Flexible plastic and electrical 15 mm components Flexible plastic and electrical 1.9 m 19 mm ClimateLine heated air components tubing SlimLine air tubing Flexible plastic 1.8 m 15 mm Standard air tubing Flexible plastic...
Flow (maximum) at set pressures The following are measured at the end of the specified air tubing: Pressure, cm H AutoSet CS-A AutoSet CS-A, AutoSet CS-A AutoSet CS-A, H5i and Standard, H5i and and SlimLine, and ClimateLine, L/min Standard, L/min...
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min. c. Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using nominal ResMed mask vent flows.
Maximum dynamic pressure variation according to EN ISO 17510-1:2009 Pressure (cm H 10 BPM 15 BPM 20 BPM AutoSet CS-A and Standard air tubing without H5i / AutoSet CS-A and Standard air tubing with H5i 0.18 / 0.18 0.30 / 0.30 0.51 / 0.51 0.21 / 0.20...
ResMed. ResMed recommends that the AutoSet CS-A device be inspected and serviced by an authorised ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the devices generally should not be required during the five year design life of the device.
Page 18
Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
Page 19
If you notice any unexplained changes in the performance of the device, if it is making unusual or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed Service Center.
Page 20
• Only use the ResMed 90W or 30W power supply unit. Use the 90W power supply unit to power the system comprising the device, H5i, air tubing, DC/DC converter and battery pack. The 30W power supply unit is designed to power the device only and recommended for travelling.