Displayed Values - ResMed AutoSet CS-A User Manual

Adaptive servo-ventilator
Hide thumbs Also See for AutoSet CS-A:
Table of Contents

Advertisement

Displayed values

Value
Pressure sensor at air outlet
Mask pressure
Flow derived values
Leak
Tidal volume
Respiratory rate
Minute ventilation
Value
Pressure measurement
Mask pressure
Flow measurements
b
Leak
b.c
Tidal volume
b,c
Respiratory rate
b, c
Minute ventilation
a. Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
b. Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL
or minute ventilation <3 L/min.
c. Measurement accuracy verified as per EN ISO 10651-6:2009 for Home Care Ventilatory Support
Devices (Figure 101 and Table 101) using nominal ResMed mask vent flows.
14
Range
4-20 cm H
4-30 cm H
0–200 L/min
0–4000 mL
0–50 BPM
0–30 L/min
Accuracy
a
±0.5 cm H
a
±12 L/min or 20% of reading, whichever is greater, at 0 to 60
L/min
±20%
±1 BPM
±20%
O (CPAP);
2
O (ASV, ASVAuto)
2
a
O (+4% of measured value)
2
Display resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min