Measuring Light Sources [Measurement Screen]; Measurement Method; Balancing Color Temperatures Of Light Sources - Konica Minolta CL-70F Operating Manual

Cri illuminance meter
Hide thumbs Also See for CL-70F:
Table of Contents

Advertisement

5.
Measuring Light Sources
[Measurement Screen]
5-1 

Measurement Method

5-1-1 

Balancing Color Temperatures of Light Sources

Take a measurement by facing the Light Receptor
measure the color temperature of the light source.
Light Receptor
1
NOTICE
● When taking measurements, the CL-70F should be located at a distance that is
approximately 10 times (10X) the diameter of the light source being measured.
● When measuring the light source, be careful not to measure the light reflected off
from ceiling, wall or floor because precesion of measurements may be affected.
● Because the precision of measurements may be affected, be careful not to
damage or dirty the white surface of the Light Receptor
becomes dirty, wipe it with a dry and soft cloth. Never use organic solvents such
as thinner or benzene.
● There may be those who are sensitive to flash or other strong lights
(photosensitivity), and they may be negatively affected by light. Therefore, be
careful when people are nearby during measurement.
5. Measuring Light Sources
towards the light source to correctly
1
. If the Light Receptor
1
60
[Measurement Screen]