MESSAGE
clean
test
area
remove
strip
5 . 6 mmol/l
(Example)
contrl
8 : 0 1am 6-01-95
error 1
retest
PROBLEM
1. There is dirt, blood, or
lint on the test area.
2. Your hand or an object
covered the test area while
turning on the meter.
The test strip was inserted
before the word
insert
appeared on the display.
If the word
appears
contrl
after a blood test, your
blood sample was too small,
smeared, or another drop was
added after the test began.
The sample was applied
before the words
apply
appeared.
sample
WHAT TO DO
1. Clean the test area
according to instructions.
2. Repeat the test. Keep test
area clear.
Remove the strip. Wait for the
word
insert
test strip.
Repeat the test with a new test
strip and apply a larger drop
of blood.
Repeat the test with a new
test strip.