Servicing Requirements - Invacare PHOENIX PLUS Service Manual

Hide thumbs Also See for PHOENIX PLUS:
Table of Contents

Advertisement

STEP 5
Loosen and detach the top crossbar link assembly using 11mm spanner and 4mm allen key
(see diagram on page 24).
STEP 6
Remove the bottom rail screws and nuts with a Philips No1 screwdriver and an
11mm spanner.
STEP 7
Remove the side frame from the crossbar assembly (see page 24)
STEP 8
Remove the crossbar bolt with a 13mm spanner, using a second spanner on the nut to hold
against rotation. Take care not to loose the washer (see page 24).
STEP 9
Remove all rubber and plastic end stops with a flat bladed screwdriver.
To reassemble reverse steps 1 to 9 remembering to lubricate the inner bottom rail with a light
grease prior to fitment.
We recommend that routine servicing be carried out at six monthly intervals. The routine ser-
vice consists of the following.
1.
Motors:
2.
Controller and Joystick Assy:
Phoenix
11/01

Servicing Requirements

a) Inspect plugs for physical damage.
b) Inspect insulation for cracks and splits.
c) Check and clean brake. Resistance = Appx. 50 - 80 Ohms.
d) Inspect and clean brushes and commutator.
e) Check action to clutch mechanism.
f) Check and inspect gear box for excess wave noise, replace grease,
recommended grease Shell Albida R2 (Q1100Z).
a) Check for physical damage to outer casings.
b) Inspect joystick shroud and gaiter for cracks and splits.
c) Check all switch operations, and Led's.
24