BATTERIES AND CHARGER
DISMANTLING - BATTERIES
STEP 1
Unplug the battery lead, disconnect the battery box from the connecting plug, lift off the
battery cradle and remove the battery cradle from the mounting bars.
STEP 2
Remove the battery boxes/battery box from cradle.
STEP 3
Unclip the battery box lids, with a 10mm spanner, undo the battery terminals and remove the
batteries from the cases.
To re-assemble, reverse steps 1 to 3 and charge the battery fully.
INSPECTION - CHARGER
STEP 1
Check the charger case for physical damage.
STEP 2
Check the input and output leads and plugs for damage, cuts and wear.
STEP 3
Check that the charging, charge complete and power on lamps are operating correctly.
STEP 4
With an ammeter or a multimeter set to amp (range 10A) test the charging current.
NOTE :
AMMETERS MUST BE CONNECTED IN SERIES WITH THE BATTERY CIRCUIT.
STEP 5
The current reading should read approximately 7amp at charge start and 0.2 amp at
complete charge.
STEP 6
With a voltmeter or a multimeter, set to volts (range 30v) test the charging voltage at the
battery supply (red + blue - ve wires). The reading should increase to 28.4v during charging.
NOTE:
NO DISMANTLING IS NECESSARY FOR CHARGERS (INSPECTION OF PLUGS ONLY).
Phoenix
12
11/01