STEP 5
Loosen and detach the top crossbar link assembly using 11mm spanner and 4mm allen key
(see diagram on page 24).
STEP 6
Remove the bottom rail screws and nuts with a Philips No1 screwdriver and an
11mm spanner.
STEP 7
Remove the side frame from the crossbar assembly (see page 24)
STEP 8
Remove the crossbar bolt with a 13mm spanner, using a second spanner on the nut to hold
against rotation. Take care not to loose the washer (see page 24).
STEP 9
Remove all rubber and plastic end stops with a flat bladed screwdriver.
To reassemble reverse steps 1 to 9 remembering to lubricate the inner bottom rail with a light
grease prior to fitment.
We recommend that routine servicing be carried out at six monthly intervals. The routine ser-
vice consists of the following.
1.
Motors:
a) Inspect plugs for physical damage.
b) Inspect insulation for cracks and splits.
c) Check and clean brake. Resistance = Appx. 50 - 80 Ohms.
d) Inspect and clean brushes and commutator.
e) Check action to clutch mechanism.
f) Check and inspect gear box for excess wave noise, replace grease,
2.
Controller and Joystick Assy:
a) Check for physical damage to outer casings.
b) Inspect joystick shroud and gaiter for cracks and splits.
c) Check all switch operations, and Led's.
Phoenix
00/04
.

Servicing Requirements

recommended grease Shell Albida R2 (Q1100Z).