STEP 2
Measure the battery voltages, the values must be the same +/- 0.2v. If not replace the
batteries.
STEP 3
For the battery charger operate, the battery voltage must be shown above 8v each. If not the
batteries must be removed and charged with an unregulated charger until the voltage
reaches 9v. Re-connect the batteries and charge through a complete charge cycle (this may
or may not recover the batteries, if not replace the batteries).
STEP 4
On completion of charge the battery voltage must read 14.20v (+/-) at 20°C or combined
28.4v.
Phoenix
11
11/01