5.
If the backrest has an electrical recliner, refasten the motor to
the backpost.
CAUTION!
Injury hazard if the clevis pin of the motor of
the electrical recliner is not correctly secured
An incorrectly secured pin can come off and the
backrest can unexpectedly fold back, which can lead
to injuries.
– Make absolutely sure that you insert the R-clip to
secure the clevis pin correctly. The clip must fit
snugly around the pin. See the illustrations below:
6.
When inserting the battery cases, please ensure that the guide
rods (1) fit exactly in the guide channels (2) provided at the side
of the battery cases, and that they are not underneath these.
1528224-Q
7.
To do this, first place the front of the battery case on the guide
rods and then push the assembly in.
Once the wheelchair is reassembled, you must ensure
that the locking system on the rear battery case has
completely engaged. Check the correct seating of the
locking pins. Also check the power supply. To do this,
switch the wheelchair on at the remote. If the electric
wheelchair is not ready to drive, check that the battery
case has been fitted correctly.
8.
Secure the battery belt with the safety clip (1).