Product Specifications - Braun WelchAllyn ThermoScan PRO 4000 User Manual

Hide thumbs Also See for WelchAllyn ThermoScan PRO 4000:
Table of Contents

Advertisement

Available languages

Available languages

6021360_PRO4000_S4-46 Seite 13 Montag, 20. August 2007 8:46 08

Product specifications

Displayed temperature range:
Operating ambient temperature range: 50–104 °F (10–40 °C)
Display resolution:
0.1 °F or °C
Accuracy for displayed temperature range:
Long term storage ranges
Temperature:
–4 °F to 122 °F (–20 °C to 50 °C)
Humidity:
95% non condensing
Shock:
withstands drop of 3 feet
Measurement time:
3–7 seconds
Automatic «power down»: 60 seconds
Battery life: 6 months/1000 measurements
Battery type: 2 x MN 1500 or 1.5 V AA (LR6)
Thermometer dimension: 6˝ x 1.7˝ x 1.3˝ (152 mm x 44 mm x 33 mm)
Thermometer weight:
3.6 oz (100 g) without batteries
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the thermometer system [ther-
mometer with probe cover]). Full responsibility for the conformance of the product to the standard is assumed by Kaz, Inc.
Hudson NY, 12534 USA.
ASTM laboratory accuracy requirements for the thermometer only (not including a probe cover) in the display range of 96.8
°F to 102.2 °F (36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).
A clinical summary is available upon request.
This device conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers (with maximum device)
The Braun ThermoScan thermometer has been proven to be safe. To ensure the best results, please take the time to read this
manual completely and keep it handy for future reference. In the unlikely event you experience any difficulty using the Braun
ThermoScan thermometer, simply call Welch Allyn toll free at: 1-800-535-6663 (in Canada:1-800-561-8797).
This products conforms to the provisions of the EC directive 93/42/EEC
(Medical Device Directive).
68–108 °F (20–42.2 °C)
±0.4 °F (95.9–107.6 °F) (35.5 °C–42 °C)
±0.5 °F (outside this temperature range)
0297