Braun ThermoScan PRO 6000 Directions For Use Manual

Braun ThermoScan PRO 6000 Directions For Use Manual

Ear thermometer
Hide thumbs Also See for ThermoScan PRO 6000:

Advertisement

ThermoScan
®
Ear thermometer
Directions for use
PRO 6000

Advertisement

Table of Contents
loading

Summary of Contents for Braun ThermoScan PRO 6000

  • Page 1 ThermoScan ® Ear thermometer Directions for use PRO 6000...
  • Page 2 © 2015 Kaz USA, Inc. Made in Mexico This product is manufactured by Kaz USA, Inc. under license to the ‘Braun’ trademark. ‘Braun’ is a registered trademark of Braun GmbH, Kronberg, Germany. ThermoScan is a registered trademark of Helen of Troy Limited.
  • Page 3 1. Braun PRO 6000...
  • Page 4 2. Package components Braun ThermoScan PRO 6000 thermometer ® Cradle Probe covers (1 or 2 probe cover boxes, depending on model) CD containing Directions for use, Quick reference guide, Welch Allyn Service Tool software and Service Tool installation guide. 2 (AA) Duracell alkaline batteries ®...
  • Page 5 4. Screen elements Battery Full battery - indicates battery is between 100% and 70% of usable battery capacity Partial battery - indicates battery is between 70% and 30% of usable battery capacity Low battery - indicates battery is between 30% and 10% of usable battery capacity Critically low battery - battery is between 10% and 1% of usable...
  • Page 6 To help ensure accurate temperature measurements, the sensor itself is warmed to a temperature close to that of the human body. When the Braun ThermoScan is placed in the ear, it continuously monitors the infrared energy until a temperature equilibrium has been reached and an accurate measurement can be taken.
  • Page 7 5.4 Why measure in the ear? Clinical studies have shown that the ear is an excellent site for temperature measurement because temperatures taken in the ear reflect the body’s core temperature . Body temperature is regulated by the hypothalamus , which shares the same blood supply as the tympanic membrane .
  • Page 8: Body Temperature

    5.5 Body temperature Normal body temperature is a range. The following table shows that this normal range also varies by site. Therefore, measurements from different sites, even if taken at the same time, should not be directly compared. Normal ranges by site Axillary 1, 2 95.6 - 99.4°F...
  • Page 9 6. Contraindications Clinical studies have shown the Braun ThermoScan thermometer can obtain accurate temperature measurements on full term newborn, pediatric and adult patients. However, there are special situations when an ear thermometer should not be used. These include, but may not be limited to the following situations: •...
  • Page 10: Battery Installation

    7. Warnings and precautions This thermometer is for professional use only. This thermometer must only be used with genuine Braun ThermoScan probe covers. Do not use any cleaning agent other than isopropyl or ethyl alcohol for cleaning the probe lens window and probe as specified in the cleaning section of this manual.
  • Page 11 8.2 Mounting instructions (Large cradle only) Mounting hardware not included. The large cradle (2 box storage) can be mounted as an easily removable wall hanger or a permanent wall mount. All mounting must be done into a wall stud. To mount the cradle the following items are needed: •...
  • Page 12: Temperature Measurement

    9. Using the Braun PRO 6000 Temperature measurement Remove thermometer from cradle by gripping thermometer at base and pivoting up. Thermometer will turn on automatically. Probe cover icon will flash on display indicating a new probe cover is needed. Attach new probe cover pushing probe tip straight into box, then pulling thermometer out.
  • Page 13 Press and release MEASURE button. The thermometer will beep once, running dashes will appear on the display then the green ExacTemp light will flash, indicating consistent probe position. Note: Always press the MEASURE button before taking a measurement. Temperature measurement. A long beep and steady green ExacTemp light will signal the end of the measuring process.
  • Page 14 Remove used probe cover by pressing Probe Cover Ejector button To achieve accurate measurements, use a new, clean probe cover for each measurement. To take another measurement, place a new, clean probe cover on the thermometer. If no action is taken the thermometer will enter SLEEP mode after 10 seconds or upon return to the thermometer cradle or host device.
  • Page 15 10.3 Manual timer The Braun PRO 6000 includes a 60 second timer that features an audible notification and visual indicator at 0, 15, 30, 45, and 60 seconds. The timer turns off automatically 5 seconds after 60 seconds is completed. The timer can be stopped at any time by pressing the Timer button or applying a probe cover.
  • Page 16 11.2 Advanced functions The following device settings must be completed using the Welch Allyn Service Tool (See 11.3 Advanced function settings and 11.4 Service tools) Default Item Description Settings Setting PerfecTemp™ Improves reading On/Off accuracy by detecting placement of probe in ear canal Celsius only selection Disables the C/F...
  • Page 17 The charging station or compatible Welch Allyn Vital Signs device can be used with the Welch Allyn Service Tool to modify settings on the thermometer. Follow the instructions to access Braun Pro 6000 advanced settings using the Welch Allyn Service Tool. Dock the Braun PRO 6000 thermometer in the Charging Station...
  • Page 18 To restore the factory default settings click Restore Defaults. Once the desired settings are selected click the Save button to send the settings to the Braun thermometer and close the box. To close the box without changing the settings click the Cancel button.
  • Page 19 11.4 Service tools For more information, refer to the Service Tool installation guide. The service tool software and service tool installation guide are available on the CD provided with the Braun PRO 6000 thermometer or from http://www.welchallyn.com/promotions/services/servicetool.htm 11.5 Charging station for storage, charging and security function (optional) A charging station is available for the Braun PRO 6000.
  • Page 20 12. Errors and notifications Error message Situation Solution No probe cover is attached Attach new, clean probe cover. (animates ON). Used probe cover is attached Discard probe cover that is on and (animates OFF). attach new, clean probe cover if taking another temperature measurement.
  • Page 21 12. Errors and notifications Error message Situation Solution System error Wait 20 seconds until the (All icons display thermometer turns off automatically, or display is blank) then turn on again. If error persists, ... reset the thermometer by removing the batteries and putting them back in.
  • Page 22 14. Maintenance and service 14.1 Cleaning and disinfecting the probe lens window and probe Only use Braun ThermoScan disposable probe covers. ® Do not use damaged, perforated, soiled or poor fitting probe covers. Do not reuse probe covers. Dirty probe lens window = lower readings. Fingerprints, cerumen, dust and other soiling components reduce transparency of the tip and result in lower temperature measurements.
  • Page 23 With probe facing down, wipe the probe with a damp cloth or cleaning wipe moistened with isopropyl or ethyl alcohol. Gently wipe with dry, clean cotton swab or cloth immediately. Allow at least 5 minutes drying time before taking temperature. Make sure probe lens window is clean and dry prior to use.
  • Page 24 As needed, clean the thermometer body and cradle following the below directions. For additional protection, we recommend placing a new probe cover on the thermometer probe to protect this area when cleaning the body of the thermometer. Use a damp cloth or cleaning wipe with cleaning solution from the Approved Cleaning Solutions Table.
  • Page 25 14.3 Cleaning and disinfecting tether (sold separately) To clean the tether, ensure that the wipe is moist, not saturated. Wipe the tether with a damp cloth or cleaning wipe with cleaning solution from the Approved Cleaning Solutions Table. 14.4 Installing new probe cover box Keep out of reach of children.
  • Page 26 14.5 Removing and installing probe cover box carrier Remove thermometer from cradle by gripping thermometer at base and pivoting up. Remove probe cover box carrier from the cradle by rotating the probe cover box carrier forward. Place probe cover box carrier back in cradle by lining up the brackets and pushing down.
  • Page 27: Replacing The Batteries

    14.7 Replacing the batteries The thermometer is supplied with two 1.5 V batteries type AA (LR 6). For best performance, we recommend Duracell alkaline batteries. ® Insert new batteries when the battery symbol begins to flash on the display (See 12. Errors and notifications). Open battery door by sliding spring latch to the right using your finger or a pointed object such as a pen.
  • Page 28 15. Specifications Displayed temperature range: 68–108°F (20–42.2°C) Operating ambient temperature range: 50–104°F (10–40°C) Display resolution 0.1°F or °C Accuracy for displayed temperature range: ±0.4°F ( (±0.2°C) (35.0°C-42°C) (95°F - 107.6°F) ±0.5°F (±0.3°C) (outside this temperature range) Clinical repeatability: <0.57°F (0.3°C) Long term storage ranges Storage temperature: -13–131°F (-25–55°C)
  • Page 29 EN 12470-5: 2003 Clinical thermometers – Part 5: Performance of infra-red ear thermometers (with maximum device) Meets CB Scheme EN 60601-1-2 :2007, IEC 60601-1-2:2007 General requirements for basic safety and essential performance—Collateral Standard: Electromagnetic Compatibility–Requirement & Tests Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and test IEC 62304 :2006 Medical device software- Software life cycle processes IEC 62366 :2007 (IEC60601-1-6) Medical devices -Application of usability engineering to...
  • Page 30 16. Warranty For Model PRO 6000 The Braun ThermoScan PRO 6000 thermometer has a limited three year warranty and the PRO 6000 cradle has a limited one year warranty.. The PRO 6000 instrument is warranted by Kaz to be free from defects in material and workmanship under normal use and service for a period of three years from the date of delivery to the first professional user who purchases the instrument.
  • Page 31 Guidance and manufacturers declaration- electromagnetic Immunity The PRO 6000 is intended for use in the electromagnetic environment specified below. The customer or the user of the PRO 6000 should ensure that it is used in such an environment. Immunity IEC60601 Compliance Electromagnetic environment test...
  • Page 32 Guidance and manufacturer’s declaration – electromagnetic emissions The PRO 6000 is intended for use in the electromagnetic environment specified below. The customer or the user of the PRO 6000 should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - Guidance The PRO 6000 uses RF energy only for it’s...