3.2 Advanced Settings/Passcode
The IQ Force dock may be set to verify a specific 4-digit passcode (PIN)
in any instrument that interfaces with the dock. If the dock detects a PIN in
the instrument that matches its own PIN, it will process the instrument. If
the dock fails to detect a PIN, or recognizes a non-matching PIN, it will not
process the instrument.
To use the passcode feature, a PIN must be entered into the IQ Force dock
Software and then uploaded to every instrument that will interface with the
dock. Each dock must then be programmed with the same PIN and set to
check for the code in the instruments.
1.
Select the dock that will be used to write the PIN (Personal
Identification Number) to the instrument. Connect the dock to the
PC either by USB port or by Ethernet port. If using an Ethernet
connection, the Ethernet setting must be configured to Live. If more
than one dock will be used to update the instruments, perform steps
4–10 for each dock.
2.
Open the IQ Force software.
32
3.
Click Advanced >> in the row of buttons on the lower right. The
software will automatically show the Security Login screen and
prompt you for your user name and password. If the Advanced button
is not shown, click on Change Config first to access it.
4.
Enter a valid User ID and Password. The Access PIN settings will
be shown:
IQ Force™ Dock Reference Manual