Connecting the Calibration Gas Cylinder to the Dock(s)
NOTE
Do not exceed the output capacity of your demand-flow
regula tor. Each IQ Force dock is able to draw calibration gas
at a maximum rate of 0.50 liters/minute. The demand flow
regulator supplied by Honeywell Analytics has a maximum
output capacity of 3 liters per minute, so it can be used to
calibrate up to four docks simultaneously. Do not connect
more than four docks to a single cylinder of calibration
gas while using the demand-flow regulator.
The 54-46-115 manifold is composed of:
• (1) small black elbow fitting
• (3) small black T-fittings
• (2) pieces of gas tubing 5" long
• (1) piece of gas tubing 10" long
• (1) piece of gas tubing 1" long
1.
Insert the demand-flow regulator into the calibration gas cylinder.
2.
Modify the tubing assembly as follows depending on how many
docks will be connected. If four docks will be used, no modifications
are necessary. For two or three docks, begin by separating the tubing
at the location described below.
12
Figure 2.
Once this is accomplished, set the two parts aside. The section that will be
used is shown in the following images (depending on the number of docks).
Figure 4. Manifold for two docks
Figure 5. Manifold for three docks
Figure 6. Manifold for four docks
Figure 3.
IQ Force™ Dock Reference Manual