Connecting the Calibration Gas Cylinder to the Dock(s)

2.6.2 PC – Multi Dock via USB
For multi-dock configurations that will be connected via USB, Honeywell
Analytics recommends the use of a powered USB hub, which is able to
power multiple USB devices directly through the USB cable. With a powered
USB hub, it is not necessary to use the wall cube power supplies that are
included with the IQ Force docks.
NOTE
With a USB hub, USB cable length from hub to dock may
not exceed 10 feet.
If you prefer to not use a powered USB hub, the following other
options are available:
If you have enough open USB ports on your PC to connect each of the
docks directly to its own USB Port, you may connect each dock directly
to the PC. Most users decide against this option either because their PC
doesn't have enough USB ports, or because it is an inappropriate use of
their PC's system resources
CAUTION
.
Under no circumstances can multiple IQ Force docks be run from
a single USB port without supplemental power.
10
2.7 Connecting the Calibration Gas Cylinder
to the Dock(s)
Connection requirements for calibration gas cylinders vary with how many
docks will be connected to the cylinder.
For instructions on connecting the gas cylinder to a single dock proceed
to
Section
2.7.1.
For instructions on connecting the gas cylinder to multiple docks proceed
to
Section
2.7.2.