Lcd Contrast Control; Battery; Calibration Interval - Honeywell IQ Force Dock Reference Manual

Dock
Table of Contents

Advertisement

Dock Controls and Instrument Status

3.1.7 LCD Contrast Control

At the top right side of the Dock Configuration box is the LCD (Liquid Crystal
Display) Contrast Control.
To darken or lighten the display on the dock, move the slider as appropriate.
The dock will not show the new LCD setting until the Set Config button
is pressed.

3.1.8 Battery

The battery level shown on the dock tab is for the internal battery in the IQ
Force dock. This battery serves to keep the real-time clock accurate while
the dock is unplugged and should not be a concern to the user unless the
battery level drops below 2.50 volts. If this happens, return the dock to
Honeywell Analytics for repair.
27

3.1.9 Calibration Interval

The dock's calibration interval settings are controlled from the Dock tab
and are located at the bottom of Hardware Configuration box. The dock's
calibration interval is the maximum number of days that the dock will allow to
pass between calibrations for any instrument that is placed in the dock. If an
instrument is placed in the dock and the interval has been exceeded, the dock
will automatically initiate a full span calibration of the instrument regardless
of the results of the bump test or the instrument's own calibration status.
To use the setting, click on Enable Cal Interval and set the interval to the
number of days between calibrations. The setting is typically enabled when
the dock is used in standalone mode.
NOTE
This setting may cause the instrument to undergo a full
calibration even when the instrument itself is not due
for calibration.
IQ Force™ Dock Reference Manual