Installing The Front Shroud/Battery Retention Bracket And Rear Shroud - Invacare TDX SI User Manual

Tdx si series power wheelchair base
Hide thumbs Also See for TDX SI:
Table of Contents

Advertisement

Installing the Front Shroud/Battery Retention Bracket and Rear Shroud

CAUTION
Wheelchairs with TRRO or TRBKTS Only - Battery retention brackets MUST be installed at all times. Otherwise, the
wheelchair will not be WC/19 compliant.
After installing front shroud, ensure that the mounting screws on the front shroud/battery retention bracket are fully
engaged into the battery box.
For this procedure, refer to FIGURE 1 on page 83.
1.
Reinstall the front shroud/battery retention bracket onto front of wheelchair frame.
2.
Install two mounting screws on the sides of the front shroud/battery retention bracket.
CAUTION
When installing batteries, ensure battery connector is securely engaged to the controller connector - otherwise serious
personal injury may result.
3.
Wheelchairs with Footboards Only: Using the mounting screw and washer, secure the footboard support to the wheelchair frame.
4.
Connect the controller to the batteries at the rear of the wheelchair.
5.
Reinstall the rear shroud and secure in place with three mounting screws.
New Batteries MUST be fully charged before using, otherwise the life of the battery(ies) will be reduced.
6.
If necessary, charge the batteries. Refer to Charging Batteries on page 56.
Part No 1154294
87
7 SETUP/MAINTENANCE
Invacare® TDX®SI Series

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tdx si-2Tdx si-hd

Table of Contents