8 TROUBLESHOOTING
NUMBER OF
ERROR CODE
FLASHES
DESCRIPTION
7
Remote Fault
8
Controller Fault
9
Communications Fault
10
General Fault
11
Incompatible or
incorrect Remote

CMPJ+, PSR+, PSF+ Joysticks or Displays

SYMPTOM
LEFT BRAKE FAULT or RIGHT BRAKE
FAULT displays and wheelchair does not
drive.
CHARGER PLUGGED IN displays.
BATTERY FAULT displays and the wheelchair
does not drive.
JOYSTICK TIMEOUT displays and the
wheelchair does not drive.
Invacare® TDX®SI Series
Check to make sure joystick is connected properly. Turn Joystick off then on.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Check joystick cable connections. Check joystick cable and connectors for damage.
Contact Invacare/Dealer for service.
Contact Invacare/Dealer for service.
Wrong type of remote connected. Contact Invacare/Dealer for service.
PROBABLE CAUSE
Motor lock levers disengaged (Error
code E9 or E10).
Battery charger connected (Error
code E28).
Batteries need to be charged (Error
code E14).
Joystick or input device is
disconnected (Error code 32).
100
POSSIBLE SOLUTION
SOLUTIONS
Engage motor lock levers. Refer to Disengaging/Engaging
Motor Lock Levers on page 59.
Unplug battery charger from the wheelchair. Refer to
Charging Batteries on page 56.
Charge batteries. Refer to Charging Batteries on page 56. If
batteries fail to charge properly, check battery charger or
replace batteries. Refer to Replacing Batteries and/or Battery
Cables on page 88.
Turn off power, reconnect the joystick of input device and
turn power on.
Part No 1154294