23.12 VSI Hand Held Programmer Diagnostic Codes

The following table should be used in conjunction with the Penny and Giles Hand Held Program
unit.
You should also refer to Penny and Giles Publication SK73747/9 for further information.
Fault Code
0A00
1500
1501
1502
1600
1E03
1E04
1E05
2C00
2F00
3000
3B00
3C00
3D00
3D01
3E00
3E01
7163
7164
716B
716C
716E
716F
7170
7173
7174
7175
7A00
7A02
7A03
All other
Codes
Trip Type
Notes
Blink
Sleep Mode. Turn controller on and off again to clear
9
General Solenoid brake Fault.
9
Short Circuit in Solenoid Brake Circuit
9
Open Circuit in Solenoid Brake Circuit
10
High Battery Voltage Fault.
Charging
Off board charger is connected, causing a drive inhibit
fault
6
Inhibit 2 is Active. Check for program error or wiring /
switch connection.
6
Inhibit 3 is Active. Check for program error or wiring /
switch connection
1
Low Battery Voltage. Voltage is below 16V
User
Joystick Displaced at power up. Allow joystick to re-
center then switch off and on again to clear.
7
Dual Joystick Displaced at power up. Allow joystick to
re-center then switch off and on again to clear.
2
Left Hand Motor Disconnected.
4
Right Hand Motor Disconnected.
3
Left Hand Motor wiring fault. Possibly short circuit
related.
3
Left Hand Motor wiring fault. Possibly short circuit
related
5
Right Hand Motor wiring fault. Possibly short circuit
related.
5
Right Hand Motor wiring fault. Possibly short circuit
related
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Dual Attendant Module Trip. The unit is defective.
8
Actuator Motor is disconnected. Check wiring.
8
Actuator Motor Wiring Trip. Possible short circuit.
8
Actuator Motor Wiring Trip. Possible short circuit.
7 or 8
Joystick or Power section Trip / Fault. Replace
Joystick.