Inspection; Disassembly / Assembly; Footplates; Legrest Assemblies - Invacare Spectra Blitz Service Instructions Manual

Hide thumbs Also See for Spectra Blitz:
Table of Contents

Advertisement

SERVICE MANUAL
Requirements:
· No 2 cross point screwdriver
· hex wrench (4mm)
· 10 mm open ended spanner
· Mallet
11.1

Inspection

· Check that the footplate fits securely and operates correctly.
· Check the operation of the footplate extension, clamping and adjustment screw (12).
· Check the legrest for damage.
11.2

Disassembly / Assembly

· Operate the legrest release mechanism and remove from the wheelchair.
· Release the height adjustment screws (13) and slide the threaded insert (14) from the hanger.
Use a mallet to gently tap the footplate if the insert will not move.

11.2.1 Footplates

· Remove screw (12) using a 4mm hex wrench.
· Separate the footplate (11) from the footplate bracket (9).
· Take care not to damage the spacers (10).

11.2.2 Legrest Assemblies

· Remove all caps (2).
· Remove fixing screws (3) and nuts using a 4mm hex wrench and a10 mm spanner.
· Remove 8 mm stud (4) and domed nut (7).
To re-assemble reverse the above steps taking care not to over tighten the fixing screws.
22
®
Invacare
- Spectra Blitz

Advertisement