General Trouble Shooting - Invacare Spectra Blitz Service Instructions Manual

Hide thumbs Also See for Spectra Blitz:
Table of Contents

Advertisement

SERVICE MANUAL
Wheelchair runs
too slowly
Remote
Power module
defective
Batteries
Electric Actuator
will not run
Actuator
Power module
/ CLAM
Remote
23.2

General Trouble Shooting

Symptom
ON/OFF button pressed to ON,
Joystick lights do not light up wheelchair
does not move.
ON/OFF lights up, wheelchair does not
move or behaves erratically.
Battery Charge indicator shows low
battery. Wheelchair moves slowly and
does not respond well to joystick
movement.
60
Inspect
Actuator and
Check / Action
· Are the battery cables connected?
· Is the joystick cable connected?
· Are the batteries disconnected?
· Are the batteries completely flat?
· Is the fuse / circuit breaker open?
· Are one or both motors disengaged
· Is the remote in Lock Mode? (See user
· Are the batteries discharged to the critical
Invacare
Replace
Power module
defective
Check /
replace
Check
Actuator wiring
Check for
physical
Replace
Actuator if
Power module
defective
Replace
remote. Check
(freewheel mode)?
guide).
level?
®
- Spectra Blitz
Section 17
Section 17
Section 14.5
Section 8.3
Section 8.3
Section 8.3
Section 17
Section 17