Control Output
Voltage (DAC) Output The multifunction module has two analog
outputs capable of outputting calibrated voltages between 12 volts with
16 bits of resolution. Each DAC (Digital-to-Analog Converter) channel can
be used as a programmable voltage source for analog input control of
other devices. A simplified diagram is shown below.
16
DAC 1
16
DAC 2
You can set the output voltage to any value between +12 Vdc and
-12 Vdc, in 1 mV steps. Each DAC is earth referenced, it cannot float.
Each DAC channel is capable of supplying 10 mA maximum current.
Note: You must limit the output current to 40 mA total for all three
slots (six DAC channels).