Chapter 8 Tutorial
Measurement Fundamentals
The 4-wire ohms method is used in systems where lead resistances can
become quite large and variable and in automated test applications
where cable lengths can be quite long. The 4-wire ohms method has the
obvious disadvantage of requiring twice as many switches and twice as
many wires as the 2-wire method. The 4-wire ohms method is used
almost exclusively for measuring lower resistance values in any
application, especially for values less than 10 and for high-accuracy
requirements such as RTD temperature transducers.
V
meter
R
I
test
370
HI-Source
HI-Sense
I
test
V
meter
LO-Sense
LO-Source