How To Use Calibration Compensation Function - Sekonic L-608 Operating Manual

Exposure meter
Hide thumbs Also See for L-608:
Table of Contents

Advertisement

6. Advanced Functions
7.

How to use Calibration compensation function

Calibration compensation can be made in precise 1/10 step increments in a +/- 1.0 EV. It may be
desired to match specific requirements, calibration to other meters, etc.
1.
Set the measurement mode (incident light, reflected light) for the desired compensation. You
can make calibration compensation independently for both incident, and reflected light. It is not
possible to switch between measurement modes if the setting is not completed.
2.
To enter the calibration setting of the meter it must first be turned off. Press the power button on
while holding down the ISO1 and ISO2 buttons simultaneously ; the screen will display CAL 0.0
(for calibration).
3.
The calibration setting can be changed by rotating the Jog wheel while pressing and holding
down the ISO 1 and ISO 2 buttons simultaneously. A range of +/- 1.0 EV in 1/10 step increments
is possible for calibration.
NOTE:
• When making calibration compensation, be sure that it satisfies your needs based
on the results of adequate test film.
• While incident and reflected light can be set independently, be aware that both
ambient light and flash exposure are corrected uniformly.
Reference:
• The calibration setting is not displayed on the main screen once it is set.