Adjusting The Axle Tube - Invacare top end terminator series User Manual

Top end terminator series
Hide thumbs Also See for top end terminator series:
Table of Contents

Advertisement

– Axle bolt C removed from axle tube D for clarity.
It is not necessary to remove the axle bolt C for
this procedure.
Adjustable Camber System
Open the camber clamp A.
1.
WARNING!
– NEVER position the camber inserts B in the
axle tube with more than 3 inches (7.62 cm) (12
indexing marks showing) of the camber insert B
outside of the axle tube D. The camber inserts
B will not be securely tightened in the axle tube
D. Possible injury to the user or damage to the
wheelchair may occur.
Position camber insert B to the desired position. Make sure
2.
there are no more than 3 inches (7.62 cm) (12 indexing marks C
showing) of the camber inserts B outside of the axle tube D.
Slide the indexing ring Eon the camber insert B until it is flush
3.
with the camber clamp A.
1171801-A
– Before using the wheelchair, make sure both
camber inserts B are set at the same indexing
mark C. This will make sure the distance between
the rear wheel and the wheelchair is the same
on both sides.
Open the camber clamp A.
4.
5.
Repeat STEPS 1–4 for opposite side of the wheelchair.

4.16 Adjusting the Axle Tube

WARNING!
– QUICK RELEASE LEVERS - Make sure the quick
release levers are in the closed position before using
the wheelchair, otherwise injury or damage to the
wheelchair may occur.
– STANDARD and suspension camber clamps - Make
sure the hex screws are securely tightened before
using the wheelchair, otherwise injury or damage to
the wheelchair may occur.
CAUTION!
– Adjustable Camber System ONLY - DO NOT close
the quick-release levers or tighten the hex screws
without camber inserts in the axle tube. Damage to
the axle tube will occur.
– Stand behind the wheelchair to determine left or right.