Paging Systems Specifications; Espa 4.4.4. Protocol - Bosch NurseCall Main Unit User Manual

Table of Contents

Advertisement

NurseCall Main Unit
A.5

Paging systems specifications

The NurseCall system can be used with the following protocols: ESPA 4.4.4, POCSAG,
Medicall 800 and DeTeWe. The following alarms or messages are sent to all systems:
For all these alarms, an acknowledgement message related to the generated alarm is sent.
A.5.1

ESPA 4.4.4. protocol

Characteristics
Only alarms arriving to the alarm buffer are transmitted, except "Relay Off". Acknowledgement
messages are also transmitted. The transmission is fully alpha-numeric. Alarms are repeated
every 1 or 3 minutes until acknowledgement.
ESPA 4.4.4. with RPE 670 / i-page
Alarms are sent to the user number attributed to the transmitter.
The default group is 00.
Group
00
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
Bosch Security Systems
Assistance
Fire
Help
Dementia
Technical
Error
Low battery
Reserve
Data rate: 9600 Bauds
Transmission: asynchronous
10 bit-structure (1 start bit, 7 data bits with even parity, 1 stop bit)
Half-duplex mode.
User number
12GG
13GG
14GG
15GG
16GG
17GG
18GG
19GG
20GG
21GG
22GG
23GG
24GG
25GG
26GG
27GG
28GG
User Manual