Differential Connections for Ground-Referenced
Signal Sources
Figure 4-6 shows how to connect a ground-referenced signal source to
an AT E Series board channel configured in DIFF input mode.
ACH<0..7>
ACH<8..15>
Input Multiplexers
AISENSE
AIGND
Selected Channel in DIFF Configuration
Figure 4-6. Differential Input Connections for Ground-Referenced Signals
With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the
signal source and the AT E Series board ground, shown as V
Figure 4-6.