INJECTING YOUR INSULIN
b
d
e
f
How to prime your InDuo™ insulin doser
• Make sure that a NovoFine
(see page 41), the push button is pressed in
completely, and the slide is closed. If the slide is
c
not closed, the release button will be locked.
• Press the release button b. You will see that the
push button jumps out c.
• Dial 1 unit d (When inserting a new Penfill
with 8 units).
• Hold the InDuo™ insulin doser with the needle
pointing upwards and tap the slide gently with
your finger a few times e.
• Press the push button completely in until it locks f.
• A drop of insulin should appear at the needle-tip.
If not, repeat the procedure until a drop of insulin
appears. If the slide is opened and closed while it
contains a partially filled Penfill
dial a larger number of units, depending on the
amount of insulin left in Penfill
Remember that priming is not finished until a drop
of insulin appears at the needle-tip.
®