The Induo™ Insulin Doser - Lifescan In Duo Manual

Combined blood glucose monitoring and insulin dosing system
Table of Contents

Advertisement

82
DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
The InDuo™ insulin doser
As with the meter display, a number of messages and symbols appear on the
InDuo™ insulin doser display during use of the doser.
Message
What it means
The very first time the insulin doser is
removed from the meter cap, the
display stays blank until the slide is
opened. However, if the display goes
off having already come on, carry out
the correction measures described in
the next column. If the display still
remains off, the insulin doser is not
working. Do not use your insulin
doser, instead call the InDuo™
Customer Care Line at 1 888 729-2299
When the insulin doser is removed
from the meter cap, all the display
segments are turned on, telling you
that the display is fully functional
(display test)
Action
If the display goes off, press
the release button, and the
push button will jump out.
Press the push button
completely in and the display
should reappear.
If the display does not reappear,
open and close the slide. Prime
the InDuo™ insulin doser as
described on pages 24 and 42
No action required