Triggering on Edges

If the signal is unstable or has a very low frequency, use
edge triggering to obtain full manual trigger control.
To trigger on rising edges of the input A waveform, do the
following:
1
Display the
TRIGGER
2
Open the
F4
3
Open the
ENTER
Shop for Fluke products online at:
key labels.
TRIGGER
menu.
TRIGGER OPTIONS
menu.
TRIGGER ON EDGE
MyFlukeStore
www.
Triggering on Waveforms
When Free Run is selected, the test tool updates the
screen even if there are no triggers. A trace always
appears on the screen.
When On Trigger is selected, the test tool needs a trigger
to display a waveform. Use this mode if you want to
update the screen only when valid triggers occur.
When Single Shot is selected, the test tool waits for a
trigger. After receiving a trigger, the waveform is displayed
and the instrument is set to HOLD.
In most cases it is advised to use the Free Run mode:
4
Select Free Run, jump to Trigger
Filter.
ENTER
5
Set Trigger Filter to Off.
ENTER
Observe that the key labels at the bottom of the screen
have adapted to allow further selection of specific edge
trigger settings: