Meter Measurements for Fluke 190-xx2

The accuracy of all measurements is within ± (% of
reading + number of counts) from 18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or
above 28 °C.
Meter Input (Banana Jacks)
Input Coupling..............................................................DC
Frequency Response.........................DC to 3 kHz (-3 dB)
Input Impedance ................ 1 MΩ (±1 %)//14 pF (±1.5 pF)
Max. Input Voltage Fluke 190-202, -102, -062:
...................................................................1000 V CAT III
(For detailed specifications, see "Safety")
Max. Input Voltage Fluke 190-502:
....................................................................1000 V CAT II
(For detailed specifications, see "Safety")
Meter Functions
Ranging ....................................................... Auto, Manual
Modes.....................................................Normal, Relative
118
600 V CAT IV
600 V CAT III
www.
Shop for Fluke products online at:
General
DC Common Mode Rejection (CMRR) ..................>100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB
Ω
Ohms (
)
Ranges..................................500.0 Ω, 5.000 kΩ, 50.00 kΩ,
Full Scale Reading
500 Ω to 5 MΩ...............................................5000 counts
30 MΩ ...........................................................3000 counts
Accuracy ............................................... ±(0.6 % +5 counts)
Measurement Current ................... 0.5 mA to 50 nA, ±20 %
Open Circuit Voltage ....................................................<4 V
Continuity (CONT)
Beep ........................................................... <50 Ω (±30 Ω)
Measurement Current ..................................0.5 mA, ±20 %
Detection of shorts of ................................................ ≥1 ms
Diode
Maximum Voltage Reading ......................................2.8 V
Open Circuit Voltage .................................................<4 V
Accuracy ............................................... ±(2 % +5 counts)
Measurement Current ...............................0.5 mA, ±20 %