Video Trigger
Standards ....PAL, PAL+, NTSC, SECAM, Non-interlaced
Modes ...................... Lines, Line Select, Field 1 or Field 2
Source ........................................................................... A
Polarity................................................. Positive, Negative
Sensitivity....................................... 0.7 division sync level
Pulse Width Trigger
Screen Update .............................On Trigger, Single Shot
Trigger Conditions............<T, >T, =T (±10 %), ≠T(±10 %)
Source ........................................................................... A
Polarity..................................... Positive or negative pulse
Pulse Time Adjustment Range........... 0.01 div. to 655 div.
with a minimum of 300 ns (<T, >T) or 500 ns (=T, ≠T),
and a resolution of 0.01 div. with a minimum of 50 ns
Shop for Fluke products online at:
a maximum of 10 s,
MyFlukeStore
www.
Continuous Auto Set
Autoranging attenuators and time base, automatic
Connect-and-View™ triggering with automatic source
selection.
Modes
Normal.................................... 15 Hz to max. bandwidth
Low Frequency ......................... 1 Hz to max. bandwidth
Minimum Amplitude A, B, C, D
DC to 1 MHz........................................................ 10 mV
1 MHz to max. bandwidth .................................... 20 mV
Automatic Capturing Scope Screens
Capacity ............................................ 100 Scope Screens
For viewing screens, see Replay function.