Symbols

Function / Meaning
Turns the device on and off.
Guide to install battery
Serial number
SN
Date of manufacture
Type BF: Device, cuff and tubing
are designed to provide special
protection against electrical
shocks.
The indicator that appears while
measurement is in progress. It
flashes when the pulse is detected.
Irregular Heartbeat indicator. (I.H.B.)
The indicator that appears when an
irregular heartbeat or any excessive
body movement is detected during
the measurement.
Flashes when there is air
remaining in the cuff at power up.
Flashes at the completion of the
measurement until exhaust is
complete.
Flashes when the pressure is not
enough for the measurement.
The battery power indicator during
Full Battery
measurement.
The battery is low when it blinks.
low Battery
Unstable blood pressure due to
movement during the
measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during inflation.
The cuff is not fastened correctly.
The pulse is not detected correctly.
EC directive medical device label
WEEE label
Symbols
ENGLISH 4
Recommended Action
Measurement is in
progress -- remain still.
Press the exhaust valve
button to exhaust air from
the cuff.
Pump air into the cuff with
the rubber bulb.
Replace the battery with a
new one, when the
indicator blinks.
Try the measuring again.
Remain very still during the
measurement.
Fasten the cuff correctly,
and try the measurement
again.