Agilent Technologies U2751A User's And Service Manual page 39

Modular switch matrix
Table of Contents

Advertisement

U2751A User's and Service Guide
SCPI Commands
The following examples show the SCPI commands for
executing the closing and opening of the relays.
Example 1, Make contact at channel 302
-> *CLS; *RST
-> ROUTe:CLOSe (@302) // Closes the relay at row 3,
Example 2, Break contact at channel 302
-> ROUTe:OPEN (@302)
Example 3, Make contact at channel 101, 302
-> ROUTe:CLOSe (@101,302) // Closes relays at row
Operation and Features
// Resets the switch to the
default power-on state.
This command can be
ignored if this operation is
not required.
column 2.
// Opens the relay at row 3,
column 2.