Operation
(AR), it will be calibrated in the range set in this menu when auto
calibration is performed. From this menu you can enable or disable the
auto calibration feature.
To navigate to the Auto Calibration Components/Range menu:
1. From the measurement mode, press the MODE key to
display the user mode.
2. Point the cursor to "Calibration Parameters" by pressing
the UP or DOWN key and press the ENT key.
3. In the "Calibration Parameters" screen that appears, point
the cursor to "Auto Calibration Components/Range" by
pressing the UP or DOWN key. Press the ENT key.
4. In the "Calibration Parameters" screen that appears, point
the cursor to "Auto Calibration Components / Range" by
pressing the UP or DOWN key. Then press the ENT key.
To set the range used for auto calibration:
1. In the "Auto Calibration Components / Range" selection
screen that appears, point the cursor to the channel whose
Teledyne Analytical Instruments
Model 7600
52