Infrared gas analyzer communication functions (modbus)
Hide thumbs Also See for 7500:
Table of Contents

Advertisement

(b) Measurement concentration (register No. 30001 to 30036)
The decimal point position and measurement unit for each concentration are stored in
registers following that of concentration, and can be known by reading them in.
The meaning of decimal point position data and measurement unit data values are the same as
in (a) above.
For example, if:
CH3 measurement concentration (register No. 30007) = 1270,
CH3 decimal point position (register No. 30008) = 2,
CH3 measurement unit (register No. 30009) = 0,
the value is 12.70 vol%
7.1.3 Handling at measurement data over-range
Even if the measurement data is at over-range, with "— — — —" displayed on the screen,
the concentration that stands then is transmitted as read-out measurement concentration.