Teledyne 7500 Instruction Manual page 22

Infrared gas analyzer communication functions (modbus)
Hide thumbs Also See for 7500:
Table of Contents

Advertisement

(2) Function explanations
Word data of continuous word number is written from write-in start address. Write-in word
data are transmitted from master station in the order of upper and lower bytes.
(3) Message transmission (example)
The following shows an example of writing the CH1 (1st component) alarm settings to No. 1
station.
CH1 range-1 high alarm setting = 1388
CH1 range-1 low alarm setting = 000A
CH1 range-2 high alarm setting = 03E8
CH1 range-2 low alarm setting = 000A
CH1 range-1 high alarm setting
Command message composition (byte)
Station No.
Function code
Write-in word number
Write-in byte number
First write-in word data
Next write-in word data
Next write-in word data
Last write-in word data
CRC data
Since the transmission data can not include a decimal point, data of 500.0 is transmitted as
"5000".
For transmission format of each data, refer to the Address map (Chapter 7).
(= 5000
H
(= 10
)
H
D
(= 1000
H
(= 10
)
H
D
Relative address: 0023
01
H
10
H
Upper
00
H
Lower
04
H
08
H
Upper
13
H
Lower
88
H
Upper
00
H
Lower
0A
H
Upper
03
H
Lower
E8
H
Upper
00
H
Lower
0A
H
Lower
E2
H
Upper
A6
H
22
Teledyne Analytical Instruments
)
D
)
D
Data number: 04
H
Response message composition (byte)
Station No.
Function code
Write-in start No.
Write-in word number
CRC data
H
01
H
10
H
Upper
00
H
Lower
23
H
Upper
00
H
Lower
04
H
Lower
30
H
Upper
00
H

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7600