Teledyne 7500 Instruction Manual page 17

Infrared gas analyzer communication functions (modbus)
Hide thumbs Also See for 7500:
Table of Contents

Advertisement

(3) Message transmission (example)
The following shows an example of reading out from No. 1 station the setting CH2 (2nd
component) range-1 zero and span calibration concentration.
Relative address of CH2 range-1 zero calibration concentration setting: 0004
Data number: 02
H
Command message composition (byte)
Station No.
Function code
Read-out start No.
(relative address)
Read-out word
number
CRC data
* Meaning of read-out data
CH2 range-1 zero calibration concentration setting 00 00
(contents of first word data)
CH2 range-1 span calibration concentration setting 03 E8
(contents of next word data)
Provided decimal point position = 1, measurement unit = 1,
CH2 range-1 zero calibration concentration = 0.0 ppm
CH2 range-1 span calibration concentration =100.0 ppm
*For handling of the decimal point and unit, refer to Section 7.1.
01
H
03
H
Upper
00
H
Lower
04
H
Upper
00
H
Lower
02
H
Lower
85
H
Upper
CA
H
Teledyne Analytical Instruments
Response message composition (byte)
Station No.
Function code
Read-out byte number
Contents of the
first word data
Contents of the
next word data
CRC data
= 0
H
= 1000
H
17
H
01
H
03
H
04
H
Upper
00
H
Lower
00
H
Upper
03
H
Lower
E8
H
Lower
FA
H
Upper
8D
H