Supported vs. Generic Transmitters

The calibrator communicates with virtually all HART
transmitters. In addition, the calibrator is programmed to
use device-dependent commands for a selection of HART
transmitters and their software versions. These are
"supported transmitters." All other transmitters are
"generic."
Table 1 shows the operations that are available for
supported vs. generic transmitters in single point and
multidrop configurations.
Sensor trim is provided for supported
transmitters, with a few exceptions, as identified
in the list of supported transmitters that you can
view on the display.
To display a list of supported transmitters and software
versions:
1.
If the calibrator is connected to a HART transmitter
with the Active Device screen showing, press the
softkey, followed by
Abort
6
Note
.
More Choices
2.
If the calibrator is not connected to a HART
transmitter, press the r key followed by
.
Choices
3.
Press the
Device Revs
appears.
Press u or d to highlight the desired manufacturer,
4.
and press e. A list of model numbers appears.
Press u or d to highlight the model number, and
5.
press e. A list of software versions appears.
More
softkey. The Browser screen
pe07s.bmp