Cloning a Transmitter

Cloning copies the Basic Setup information from one
transmitter to another. You can clone generic and
supported transmitters.
To use the calibrator to clone a transmitter:
1.
Connect the calibrator to the transmitter you want to
clone. You only need a communication connection.
2.
From the Active Device screen, press the
softkey.
Use the u or d keys to select
3.
the screen and press e.
Setup
from the list on
Basic
pe09s.bmp
Documenting Process Calibrator HART Mode
4.
Press the
softkey to load the settings into
Store Page
the calibrator's memory (not the transmitter).
5.
Disconnect the calibrator from the transmitter and
connect it to the transmitter you want to configure.
Press the
softkey to establish communication.
Abort
6.
Press the
softkey, then select
Setup
the Basic Setup screen.
7.
Press the
Recall Page
display with the cloned parameter settings. The
settings are not transmitted to the transmitter at this
time.
Use the u or d and ekeys to select
8.
parameters one at a time, change them, or leave them
as cloned, and press the
them to the transmitter.
Cloning a Transmitter
to return to
Basic
softkey to refresh the calibrator
softkey to transmit
Send
19