Front Panel Disassembly; Reassembly Procedure - Fluke 8842A Instruction Manual

Hide thumbs Also See for 8842A:
Table of Contents

Advertisement

8842A
Instruction Manual

6-32. Front Panel Disassembly

1. Holding the chassis vertically (with the front panel downward), remove the mounting
2. Holding the chassis vertically (now standing the instrument on the rear panel), pull
3. Remove the the display assembly (Display PCA, spacer matrix, and keypad) from the
4. Separate the spacer matrix from the Display PCA by releasing the two pairs of plastic
5. Remove the keypad from the spacer matrix.
The vacuum fluorescent display should not be removed from
the Display PCA; these are supplied as one part.
6. Remove the display window from the front panel as follows:

6-33. REASSEMBLY PROCEDURE

To reassemble the instrument, proceed as follows:
1. Assemble the front panel assembly by reversing the front panel disassembly
The four Display PCA mounting screws are self-tapping. To
avoid damaging the threads, ensure the screws are threaded
properly before tightening. Do not overtighten them.
2. Turn the chassis upside down.
3. Install the Main PCA through the bottom of the chassis as follows:
6-32
screws from the four corners of the Display PCA (Figure 6-7A).
the front panel off the chassis and set it aside (Figure 6-7B).
chassis as follows (see Figure 6-7C):
a. Release the two plastic tabs on the front of the chassis.
b. Let the bottom edge of the display assembly swing toward the rear of the
instrument.
c. Pull the Display PCA toward the bottom of the chassis.
snap tabs on the back of the Display PCA (Figure 6-7D).
a. Slide the window upward (away from the buttons) about 1/32 inch (Figure 6-8A).
b. Push the window directly outward from the front panel (Figure 6-8B).
procedure (Figure 6-7). (It is easiest to lay the keypad on the Display PCA before
installing the spacer matrix on the Display PCA.)
CAUTION
CAUTION