Dc Calibration Connections; Dc Voltage Test - Fluke 8842A Instruction Manual

Hide thumbs Also See for 8842A:
Table of Contents

Advertisement

8842A
Instruction Manual
STEP
RANGE
1
A
1
B
200 mV
1
C
2V, 20V,
D
E
200 mV
F
G
H
I
1. Relative to high-quality short stored using OFFSET feature.
6-6
Figure 6-1. DC Calibration Connections
INPUT
(V dc)
MIN
20 mV
0V
-0.0030
(short)
0V
-00.003
(short)
0V
-2 counts
200V,
(short)
1000V
20 mV
10 mV
+9.9963
100 mV
+99.993
2V
1V
+.99996
20V
10V
+9.9995
200V
100V
+99.995
1000V
1000V
+999.95
Table 6-2. DC Voltage Test
Displayed Reading
SLOW
MAX
MIN
+0.0030
-0.0050
+00.003
-00.005
+2 counts
-4 counts
+10.0037
+9.9943
+100.007
+99.991
+1.00004
+.99994
+10.0005
+9.9993
+100.005
+99.993
+1000.05
+999.93
MEDIUM
MAX
MIN
+0.0050
-0.030
+00.005
-00.03
+ 4 counts
-3 counts
+10.0057
+9.970
+100.009
+99.97
+1.00006
+.9997
+10.0007
+9.997
+100.007
+99.97
+1000.07
+999.7
f6-01.wmf
FAST
MAX
+0.030
+00.03
+3
counts
+10.030
+100.03
+1.0003
+10.003
+100.03
+1000.3

Hide quick links: