Input Bias Current Error; Measuring Input Bias Current Error - Fluke 8842A Instruction Manual

Hide thumbs Also See for 8842A:
Table of Contents

Advertisement

The input impedance of the 8842A is 10 M in the 200V and 1000V dc ranges, and is
greater than 10,000 M in the 20 mV, 200 mV, 2V, and 20V ranges. Therefore, for the
8842A, circuit loading error is less than 0.01% as long as the source impedance is less
than 1 M in the 20 mV, 200 mV, 2V, and 20V ranges, and less than 1 k in the 200V
and 1000V ranges. The exceptionally high input impedance on the 20V dc range allows
high-accuracy readings in CMOS and high-impedance analog circuitry.
Input protection circuitry can reduce the input impedance to as low as 100
k when the input is overrange. This may also occur momentarily when the
instrument autoranges to a higher range.
4-4.

Input Bias Current Error

Input bias current error occurs because a voltmeter's input bias current always changes
the voltage of the circuit under test. However, the error is significant only when
measuring voltages in circuits with very high source impedance. The error can be
measured as shown in Figure 4-2.
With the 8842A, it is easy to correct for this error using the OFFSET button:
1. Select the VDC function and the desired range.
2. Connect the 8842A INPUT terminals to a resistor which matches the source
impedance of the circuit to be tested.
NOTE
Figure 4-2. Measuring Input Bias Current Error
Measurement Tutorial
DC VOLTAGE MEASUREMENT
f4-02.wmf
4
4-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents