Main Chassis Disassembly; Front Panel Disassembly; Assembly Procedures - Fluke 8845A Calibration Manual

Digital multimeter
Hide thumbs Also See for 8845A:
Table of Contents

Advertisement

8845A/8846A
Calibration Manual
6. Remove the wedges from the front and rear input modules by rotating the top toward
7. Remove both sets of screws holding the front panel to the chassis (8-32 pan head and
8. Gently remove the front panel by pulling it forward and set it aside.
9. Remove the plastic handle caps from the chassis sides by rotating the front slightly

Main Chassis Disassembly

To disassemble the Meter's main chassis:
1. Carefully remove the transformer connectors from the main board.
2. Carefully remove the screws that hold the transformer and its bracket to the chassis
3. Carefully lift the transformer out of the chassis and place the transformer behind the
4. Remove the jackscrews for the RS-232 and IEEE488 connectors to the chassis.
5. Remove the screws holding the main circuit board to the chassis (three 6-32 pan-head
6. Remove the main circuit board.

Front Panel Disassembly

To disassemble the Meter's front panel:
1. Remove the one 6-32 pan head and two 6-32 flat head undercut screws holding the
2. Remove the thread-forming screw holding the front panel shield to the front panel
3. Remove the three thread-forming screws from the keypad assembly and remove

Assembly Procedures

To assemble the three parts of the Meter, follow the disassembly instructions in reverse
order.
2-8
the chassis middle and pulling up. See Figure 4-1.
6-32 flat head undercut). See Figure 4-1.
inward and pulling forward.
while holding the transformer so that it cannot drop on the circuits.
instrument with the power module leads still attached.
screws).
display shield to the rest of the front panel assembly and remove the shield.
and remove the shield.
display module.