3.3 GPIB interface controls

The GPIB port is a dedicated interface module of UTD4000 Series Digital Storage Oscilloscopes. After
installing the communication and control software, you can carry out system operation and control your
oscilloscope through the GPIB interface. You can create an automatic testing system simply, quickly and
efficiently, and carry out testing and measuring.
With the optional UT-M06 extension module, you can operate your UTD4000 oscilloscope through the
GPIB interface as follows :
1.
Select "GPIB" in the equipment selection option on the real-time control panel.
2.
When "GPIB" is selected, the dialog box in Fig 3-6 will pop up.
Enter the GPIB device address in this box (range : 1-30). The GPIB device addresses set in the software
and the communicated oscilloscope must be the same, otherwise communication between the
oscilloscope and the software cannot be established.
Access sub-menu page 2 by pressing [UTILITY] → GPIB on the oscilloscope control panel for GPIB
device address setup.
There must not be two identical GPIB device addresses on one GPIB main line.
3.
For other operation instructions refer to steps 2 and 3 for "USB interface controls" in this chapter.
Note : When using the GPIB interface, the "transmission interface" feature is disabled.
UTD4000 Digital storage oscilloscope communication control software Operating Manual and Installation Guide
Fig 3-6
Page 24 of 27