Dc Current Output - Fluke 744 User Manual

Documenting process calibrator
Hide thumbs Also See for 744:
Table of Contents

Advertisement

744
Users Manual

DC Current Output

Range/Mode
22 mA/ Source mA
22 mA/ Simulate Transmitter
(Current Sink)
Maximum Burden Voltage: 24 V
Temperature Coefficient: (0.003% of output + 0.003% of f.s.)/ C in the ranges -10 to 18 C and 28 to 50 C
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30 V dc
Specification applies for currents between 2 mA and 22 mA. For current below 2 mA, typical accuracy is 0.15% of full
scale.
110
Resolution
1 A
0.01% + 0.015%
1 A
0.02% + 0.03%
% of Output + % of Full Scale
1 Year
2 Year
0.02% + 0.015%
0.02% + 0.03%

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents