Continuity Testing; Frequency Measurement - Fluke 744 User Manual

Documenting process calibrator
Hide thumbs Also See for 744:
Table of Contents

Advertisement

744
Users Manual

Continuity Testing

Tone
Continuous tone
May or may not get tone
No tone

Frequency Measurement

Ranges
1.00 Hz to 109.99 Hz
110.0 Hz to 1099.9 Hz
1.100 kHz to 10.999 kHz
11.00 kHz to 50.00 kHz
Minimum Amplitude for Frequency Measurement (square wave):
<1 kHz: 300 mV p-p
1 kHz to 30 kHz: 4 mV p-p
>30 kHz: 2.8 V p-p
Maximum input:
<1 kHz: 300 V rms
>1 kHz: 30 V rms
Input Impedance: 5 M
For frequency measurement less than 109.99 Hz, specifications apply for signals with a slew rate greater than 5
volt/millisecond.
108
Resistance
<25
25 to 400
>400
Accuracy
1 Year
0.05 Hz
0.5 Hz
0.005 kHz
0.05 kHz
2 Year
0.05 Hz
0.5 Hz
0.005 kHz
0.05 kHz