Thermocouple Types Accepted - Fluke 744 User Manual

Documenting process calibrator
Hide thumbs Also See for 744:
Table of Contents

Advertisement

Type
Positive Lead
Material
E
Chromel
N
Ni-Cr-Si
J
Iron
K
Chromel
T
Copper
B
Platinum (30% Rhodium)
R
Platinum (13% Rhodium)
S
Platinum (10% Rhodium)
C ***
Tungsten (5% Rhenium)
L (DIN J)
Iron
U (DIN T)
Copper
*American National Standards Institute (ANSI) device negative lead (L) is always red.
**International Electrotechnical Commission (IEC) device negative lead (L) is always white.
*** Not an ANSI designation but a Hoskins Engineering Company designation.
Table 5. Thermocouple Types Accepted
Positive Lead (H)
Color
ANSI*
IEC**
Purple
Violet
Orange
Pink
White
Black
Yellow
Green
Blue
Brown
Gray
Black
Orange
Black
Orange
White
Documenting Process Calibrator
Negative Lead
Material
Constantan
Ni-Si-Mg
Constantan
Alumel
Constantan
Platinum (6% Rhodium)
Platinum
Platinum
Tungsten (26% Rhenium)
Constantan
Constantan
Using Measure Mode
Specified Range
( C)
-250 to 1000
-200 to 1300
-210 to 1200
-270 to 1372
-250 to 400
600 to 1820
-20 to 1767
-20 to 1767
0 to 2316
-200 to 900
-200 to 600
35