Calibrating the scanner
A-61169 January 2005
4. Select Return.
Elevator position : 1000
5. Use the down arrow to select the desired elevator position.
6. Select Return. The elevator position takes effect after the scanner is
enabled.
Calibration optimizes your scanner to achieve the best image quality and
feeding performance. The scanner can only be calibrated when it is
disabled.
You should calibrate:
when a message is displayed requiring calibration
when image quality is poor
when paper handling is degraded (false multi-feeds)
after cleaning the scanner
Use the calibration target (shown below) provided with your scanner. Be
sure to use a good, clean calibration target. Additional calibration targets
can be ordered. See Chapter 5, Maintenance, "Ordering Supplies" for
more information.
To calibrate the scanner:
1. Select the Calibrate
2. Place the calibration target in the elevator tray face up, arrows first.
3. Press Start/Resume.
4. Place the calibration target in the elevator tray face down, arrows first.
5. Press Start/Resume. Calibration results will be displayed.
icon.
3-13