Previously recorded data
The Calibration program also allows you to work with
previously recorded calibration data.
If you have used the ER60 to record raw data from a previous
calibration session you may operate the ER60 in replay mode
and perform the described calibration procedure as if the ER60
was operating in Normal mode. When operating in Replay
mode you can set the ping rate to maximum since this will not
affect the data quality.
Offline calibration
The calibration program may be run offline without an
operational ER60. To start it, locate the file calibration.exe. If
you have performed a standard installation, this file is normally
found in the following path:
c:\program files\simrad\scientific\ek60\bin\calibration.exe
If you have a calibration file previously generated by the ER60
calibration program, you can import the file in the calibration
program.
1
Open the File menu in the calibration window, choose
Open.
2
Select the file you would like to import, and click OK.
3
Examine and edit the previously recorded calibration data.
Edited data may be saved and you can update the transducer
parameters using the imported or edited calibration data. This
also enables you to use different calibration files from different
environmental conditions.
Related topics
Environment, page 77
Normal Operation, page 102
Ping Control, page 105
Single Target Detection, page 117